Optimization and audit of radiation dose during percutaneous transluminal coronary angioplasty


1 Department of Radiology, Christian Medical College, Vellore, Tamil Nadu, India
2 Department of Cardiology, Christian Medical College, Vellore, Tamil Nadu, India

Correspondence Address:
Roshan S Livingstone
Department of Radiology, Christian Medical College, Vellore - 632 004, Tamil Nadu
India
Login to access the Email id


DOI: 10.4103/0971-6203.37478

PMID: 21224923

Get Permissions

 

   Abstract  

The percutaneous transluminal coronary angioplasty (PTCA) is one of the interventional procedures which impart high radiation doses to patients compared to the other cardiologic procedures. This study intends to audit and optimize radiation dose imparted to patients undergoing PTCA. Forty-four patients who underwent PTCA involving single or multiple stent placement guided under cardiovascular X-ray machine were included in the study. Radiation doses were measured using dose area product (DAP) meter for patients undergoing single and multiple stent placements during PTCA. A dose reduction of 27-47% was achieved using copper filters and optimal exposure parameters. The mean DAP values before optimization were 66.16 and 122.68 Gy cm 2 for single and multiple stent placement respectively. These values were 48.67 and 65.44 Gy cm 2 respectively after optimization. In the present scenario, due to the increase in the number of PTCAs performed and the associated risk from radiation, periodical audit of radiation doses for interventional procedures are recommended.


Keywords: Dose area product; radiation dose


How to cite this article:
Livingstone RS, Timothy Peace B S, Chandy S, George PV, Pati P. Optimization and audit of radiation dose during percutaneous transluminal coronary angioplasty. J Med Phys 2007;32:145-9

How to cite this URL:
Livingstone RS, Timothy Peace B S, Chandy S, George PV, Pati P. Optimization and audit of radiation dose during percutaneous transluminal coronary angioplasty. J Med Phys [serial online] 2007 [cited 2014 Feb 28];32:145-9. Available from: http://www.jmp.org.in/text.asp?2007/32/4/145/37478


Cardiovascular interventions are complex procedures performed with dedicated fluoroscopy machines. In recent times, these interventions are rapidly replacing sizable fraction of cardiovascular surgeries. The PTCA involves applications such as stent deployment, resulting in high radiation doses due to increasingly long exposure times. [1],[2] The duration of PTCA depends on the complexity of the disease. The more complex the anatomy, the longer the duration of fluoroscopic screening and the number of cine runs acquired. During PTCA, the radiation doses to patients are relatively high. [3],[4],[5],[6] It is therefore prudent to perform dose auditing on this procedure by implementing radiation safety protocols, ensuring good work practices and comparing risks involved with benefits obtained. Establishing reference levels for interventional examinations presents a problem since patient numbers are limited and these interventional procedures are often performed at a few specialist centers. [7]

Radiation doses imparted to patients can be measured using a dose area product (DAP) meter. The DAP is the most reliable measurement technique for dynamic examinations such as fluoroscopy in which the projection technique and the technique parameters are continually varying. [7],[8] Moreover, DAP allows the estimation of patient dose in complex examinations and does not depend on the distance from the X-ray tube. [9] The DAP is particularly useful for assessing and comparing radiation doses from screening procedures, and it provides useful indication of the overall patient exposure rather than measurement of surface dose and interprets doses to particular organs. [10] Though DAP is an ancient measurement tool, it still remains as one of the best ways to measure radiation doses from diagnostic and interventional cardiologic and radiological procedures.

Dose auditing during radiological procedures is of great value in optimizing radiation dose to patients without compromising diagnostic yield. Modification of key imaging parameters such as tube voltage, mAs, field collimation, geometric magnification and radiation exposure to the image receptor has an impact on image quality and the radiation dose to the patient and these require optimization. The present study is intended to audit and optimize radiation dose imparted to patients during PTCA procedure.


   Materials and Methods   Top


All PTCA procedures were performed using Philips Integris H5000 (the Netherlands) machine dedicated for cardiovascular procedures. The machine incorporated preprogrammed radiographic factors to suit the region of anatomical interest. It also had several selection modes with specific table-top dose rates such as low, normal and high for fluoroscopy screening. In the initial stages of dose auditing, the table-top dose rate for 'low fluoroscopy mode' was set to 43.8 mGy/min, 'normal' to 87.5 mGy/min and 'high' to 175 mGy/min. The Philips H5000 machine incorporated a minimum filtration of 1.5 mm Al equivalent; and added filtration (spectral filters) of 0.1, 0.2 and 0.4 mm copper. The spectral filters were automatically selected by the machine according to the dose rate settings selected by the operator. The machine had a Charged Couple Device (CCD) camera fixed along with the image intensifier. The advantage of using CCD camera was that it necessitated lower exposure factors, thus imparting less radiation dose to patients.

The optimization process (P2*) involved halving the dose rates to 21.9, 43.8 and 87.5 mGy/min for fluoroscopy selection modes of low, normal and high respectively. This optimization was achieved by increasing the existing filtration specified by the manufacturer and by increasing tube potentials and lowering the tube current. The dose optimization techniques using copper filters are still practiced, and more literature is required in the current scenario due to the arrival of flat panel machines. Ideally, the kVp and the filter should be preferably chosen so that for the resultant X-ray spectrum, the sensitivity of the detector or detector combination should be maximum. In addition, the energy should be optimized for better visibility of narrow blood vessels with contrast agents. For the kVp and filter used in our study, cardiologists found that the images were of optimum quality for performing the procedures.

PTCA procedure

The PTCA involves opening stenosed segments in coronary vessels using balloon angioplasty and deploying stents for long-term patency in majority of cases. After gaining femoral arterial access using the Seldinger technique, patients are taken for PTCA directly if a coronary angiogram (CA) was already available. Using dedicated guiding catheters, the coronary ostea is cannulated. After proper angiographic delineation, the optimum views are selected to show the site and length of lesion without fore shortening. The lesion is crossed with 0.014 inch sized wires and dilated with appropriate sized balloons. Adequately sized premounted stents are then deployed at the lesion site for long-term patency. Final cine runs in orthogonal views confirmed the adequacy of the procedure. The PTCA was performed by one consultant cardiologist and one resident cardiologist. The PTCA was categorized into two groups namely patients who had a single stent placement and patients who had multiple stent placement.

It is possible to use one of the three image intensifier formats (IIFs) or field sizes with diameters such as 23, 17 and 14 cm available on both the machines. A 23-cm IIF was used during fluoroscopic screening in PTCA procedure for tracing the path of the catheter from the region of femoral puncture and in the cardiac valve region. The fluoroscopic screening along the femoral region while tracing the path of the catheter from the region of femoral puncture was done only if there was a problem in the smooth insertion of the wire. A 17-cm IIF was used for the oblique, caudal, cranial and lateral projections delineating the coronary anatomy during the procedure. The 14-cm IIF was not used unless there was a necessity for finer details while performing the procedure. Selection of these IIFs was at the discretion of the personnel performing the procedure.

Dosimetry

Radiation dose imparted to patients who underwent PTCA procedure was measured using DAP meter (diamentor PTW; Freiburg, Germany) which was fitted to the collimator assembly [Figure - 1]. The DAP meter is constructed using transparent plastic, and it is therefore completely unobstructive to the examination. The DAP meter measured radiation dose contributed from fluoroscopy screening and cine runs. During the course of the examination, personnel involved in the data collection continuously monitored DAP values and this facilitated acquisition of DAP values pertaining to fluoroscopy screening and cine runs separately. Periodic calibration of the DAP meter was done by company engineers every 6 months in the presence of medical physicists.


   Results   Top


Out of the 44 patients who underwent PTCA procedure, 6 were female patients and 38 were male patients. [Table - 1] shows patient-related parameters along with the exposure factors used during fluoroscopic screening and cine runs. The exposure parameters used for single stent placement and multiple stent placement during PTCA with company preset values for the H5000 Philips Integris machine are shown as P2 in [Table - 1]. The P2* represents the optimized values used in the same machine. After the optimization, it is found that tube potentials during fluoroscopic screening were higher for P2* compared those used in P2; however, the tube current in P2* was less than that used in P2. The kVp and mA were only slightly increased for cine runs, and the tube potential was increased as for fluoroscopic screening in both P2 and P2*. The fluoroscopic time duration during the single stent placement ranged from 4.9 to 29.9 min; while for multiple stent placement, it ranged from 7.9 to 25.8 min. The PTCAs were performed under automatic brightness control, in which the tube potential and tube current were adjusted.

[Table - 2] shows the DAP values for PTCA. Though the time duration of fluoroscopic screening was higher than the time duration of cine runs, the percentage of radiation dose imparted to patients from cine runs was higher than that from fluoroscopic screening. The mean DAP value for single stent placement in P2 was 66.16 Gy cm 2 , whereas the mean DAP value after optimization (P2*) was 48.67 Gy cm 2 . Similarly for multiple stent placement, the mean DAP value in P2 was 122.68 Gy cm 2 and mean DAP value for P2* was 65.44 Gy cm 2 . The Student's ' t ' test showed that the DAP value from P2 for PTCA was significantly higher than that for P2* ( P < 0.001). [Table - 2] also shows the dose rates during cine runs and fluoroscopic screening. After the optimization (P2*), the use of high tube filtration and low tube currents has reduced the dose rates by about 40-50% for fluoroscopic screening and cine runs respectively and hence has an effect on the overall radiation dose imparted to patients during PTCA.


   Discussion   Top


The PTCA is an interventional cardiologic procedure involving high fluoroscopic doses and replaces a sizable fraction of cardiovascular surgeries in the current scenario. Repeated PTCAs may be required for some patients in order to achieve the desired result. In spite of being a procedure of imparting high radiation doses to patients, the benefits obtained from this procedure far outweigh the risk involved from radiation. The potential deleterious consequence of this procedure involves stochastic effects as well as deterministic effects. [11]

There are no standard protocols available for PTCA; hence it is necessary to audit and optimize radiation doses during this procedure. During the optimization process in the current study, adequate copper filtration with high tube potentials and low tube currents was selected. It has been observed that with suitable thickness of copper filtration combined with optimized exposure factors, absorbed dose during fluoroscopic procedure can be reduced with little or no loss of image quality. [12] As pointed out by Fenner et al ., there is a reduction of 40% of absorbed dose with only minimal loss of image quality with the using 0.2 mm Cu filter. [13] Use of tube potentials of 120-125 kV instead of 90-100 kV and corresponding mAs could reduce entrance surface doses by 50% and effective dose by 43%. [14]

In the current study, the radiation doses after optimization (P2*) were found to be 27% lower for single stent placement and 47% lower for multiple stent placement compared to those in P2 using company preset values. It is also noteworthy in this context that the mean DAP value of 65.44 Gy cm 2 for multiple stenting in P2* was comparable to the mean DAP value of 66.16 Gy cm 2 for single stenting in P2. From [Table - 2], it is also noted that dose rates for cine runs and fluoroscopy screening are reduced in P2*. Since the PTCA is usually preceded by a diagnostic CA, it would be recommended to include radiation dose from CA along with that involved in PTCA also, since radiation doses have a cumulative effect. The mean DAP value for patients undergoing CA procedure before optimization was 55.86 Gy cm 2 . In a separate study conducted earlier, the mean DAP value for CA procedure was 27.71 Gy cm 2 . [15] Adding the DAP value after the optimization for corresponding patients for both CA and PTCA would therefore give 76.38 and 93.15 Gy cm 2 for single stent and multiple stent placement respectively. No ill effects of radiation, such as deterministic or stochastic, were reported for any patient during the course of the study.

The mean DAP value after the optimization in the current study for single stent placement was 48.67 Gy cm 2 , and this lies well within the values of radiation doses reported in literature. [Table - 3] shows a comparison of DAP values with those from studies reported in literature. [4],[8],[16],[17],[18],[19],[20],[21] Kuon et al ., in their study, have optimized radiation doses up to 14.4 Gy cm 2 for PTCA by adopting techniques like training the staff in the fluoroscopy room, using low fluoroscopy modes and by limiting the number of cine runs wherever possible to one heart cycle length. [21]

High radiation doses are imparted to patients who require cranio-caudal angulations since they require maximum X-ray output and involve high tube loadings. Mottled appearance was visualized in both P2 and P2* during these extensive angulations. Short cine runs were required to visualize the vessels prominently if the images gave mottled appearance during fluoroscopic screening, and this requires increase of radiation doses. However, after the optimization, the images did not produce significant mottled appearance which could degrade the diagnostic information. In the current study, the length of the procedure and the duration of fluoroscopic screening and cine runs acquired depended upon the complexity of anatomy of the patient and the number of stents needed to be implanted. The increase in the number of cine runs and frames acquired during cine runs depended on the personnel performing the procedure and complexity of the disease. A few cardiologists required higher number of frames for each run in order to visualize the vessels more prominently. Radiation doses were high during the placement of stent when high degree of tube angulations, especially cranio-caudal positions, was selected for certain patients.

The cardiovascular machines are specifically designed for interventional cardiologic procedures and have the option of using different IIFs. Appropriate selection of IIFs based on exposure parameters, dose rates and image magnification required to obtain the necessary clinical information could reduce radiation dose imparted to the patient significantly. The cardiovascular machines dedicated for cardiac procedures are invariably imported from developed countries and the default preset exposure settings are based on the patient size of those specified locations. Since the weight and body surface area of Indian patients are lower in comparison with patients of other countries, it would be advisable to set up appropriate dose modes for the Indian population. To achieve low dose levels during cardiac procedures, the image intensifier entrance dose rates could be reduced according to the body mass index of the Indian population. However, radiation-intensive angulations do not enable substantially better image quality despite increasing image intensifier dose levels. [22]

The dose reduction achieved by these methods is likely to translate into low doses to the operators as well. Further dose reduction is possible by adopting pulsed fluoro mode operation. As pointed out by Kuon et al ., use of a 25 ps -1 would also impart low radiation dose to patients if adequate radiation safety standards are maintained. [22] The dose can be further reduced by using 12.5 ps -1 in cine mode rather than using 25 ps -1 . Maintenance and quality assurance checks at regular intervals are mandatory. Improperly calibrated machines can also inadvertently impart large doses to patients without the knowledge of the operator. [23]

The reference DAP dose levels presently available for developed countries are 45 and 75 Gy cm 2 for CA and PTCA respectively, and our center fulfils the above criteria for these interventional procedures. In accordance with the ALARA principle, interventionists should therefore vary cine image intensifier dose rates and may to a certain extent compromise optimal image quality in accordance with diagnostic and therapeutic requirements, documented structure, tube angulation and body mass index. [22]


   Conclusion   Top


Though the radiation dose imparted to patients does not present any alarming situation with regard to ill effects of radiation, it would be prudent to optimize radiation dose to patients undergoing PTCA and take efforts towards achieving reduction in radiation dose to the patients. Since there is a frequent change in the various imaging modalities, reference dose levels should be audited on a time-to-time basis so as to keep the doses as low as reasonably practical. Enhanced knowledge of 'radiation dose'-reduction techniques significantly reduces patient radiation hazards in invasive cardiology.


   Acknowledgment   Top


The authors would like to thank all the staff members in the catheterization lab for their cooperation with the team for collecting the data.

 
   References   Top

1. Mcparland BJ. Entrance skin dose estimates derived from dose-area product measurements in interventional radiological procedures. Br J Radiol 1998;71:1288-95.  Back to cited text no. 1  [PUBMED]  [FULLTEXT]
2. Driscoll DO, McNeil EA, Ferrando J, Slaney P. Effective dose to the patient undergoing superior vena cava stent. Br J Radiol 1998;71:1302-5.  Back to cited text no. 2    
3. Faulkner K. Radiation protection in interventional radiology. Br J Radiol 1997;70:325-6.  Back to cited text no. 3  [PUBMED]  [FULLTEXT]
4. Vano E, Gonzalez L, Fernandez JM, Guibelalde E. Patient dose value in interventional radiology. Br J Radiol 1995;68:1215-20.  Back to cited text no. 4    
5. Cascade PN, Peterson LE, Wajszczuk WJ, Mantel J. Radiation exposure to patients undergoing percutaneous transluminal coronary angioplasty. Am J Cardiol 1987;59:996-7.  Back to cited text no. 5  [PUBMED]  [FULLTEXT]
6. Finci L, Meier B, Steffeno G, Roy P, Rutishauser W. Radiation exposure during diagnostic catheterization and single and double-vessel percutaneous transluminal coronary angioplasty. Am J Cardiol 1987; 60:1401-3.  Back to cited text no. 6    
7. Marshall NW, Chapple CL, Kotre CJ. Diagnostic reference levels in interventional radiology. Phys Med Biol 2000;45:3833-46.  Back to cited text no. 7  [PUBMED]  [FULLTEXT]
8. Betsou S, Efstathopoulos E, Katritsis D, Faulkner K, Panayiotakis C. Patient radiation doses during cardiac catheterization procedures. Br J Radiol 1998;71:634-9.  Back to cited text no. 8    
9. Ruiz-Cruces R, Pιrez-Martνnez M, Martνn-Palanca A, Flores A, Cristófol J, Martνnez-Morillo M, et al . Patient dose in radiologically guided interventional procedures: Conventional versus digital systems. Radiology 1997;205:385-93  Back to cited text no. 9    
10. Broadhead DA Chapple CL, Faulkner K. The impact of digital imaging on patient doses during barium studies. Br J Radiol 1995;68:992-6.  Back to cited text no. 10    
11. Wagner LK, Eifel PJ, Geise RA. Potential biological effects following high x-ray dose interventional procedures. J Vasc Interv Radiol 1994;5:71-84.  Back to cited text no. 11  [PUBMED]  
12. Tapiovaara MJ, Sandborg M, Dance DR. A search for improved technique factors in paediatric fluoroscopy. Phys Med Biol 1999;44:537-59.  Back to cited text no. 12  [PUBMED]  [FULLTEXT]
13. Fenner JW, Morrison GD, Kerry J, West N. A practical demonstration of improved technique factors in paediatric fluoroscopy. Br J Radiol 2002;75:596-602.  Back to cited text no. 13  [PUBMED]  [FULLTEXT]
14. Martin CJ, Hunter S. Reduction of patient doses from barium meal and barium enema examinations through changes in equipment factors. Br J Radiol 1994;67:1196-205.  Back to cited text no. 14  [PUBMED]  
15. Livingstone RS, Chandy S, Timothy BS, George PV, Bobby J, Purendra P. Audit of radiation dose to patients during coronary angiography. Indian J Med Sci 2007;61:83-90.  Back to cited text no. 15    
16. Broadhead DA, Chapple CL, Faulkner K, Davies M, McCallum H. The impact of cardiology on collective effective dose in the North of England. Br J Radiol 1997;70:492-7.  Back to cited text no. 16    
17. Efstathopoulos EP, Makrygiannis SS, Kottou S, Karvouni E, Giazitzoglou E, Korovesis S, et al . Medical personnel and patient dosimetry during coronary angiography and intervention. Phys Med Biol 2003;48:3059-68.  Back to cited text no. 17  [PUBMED]  [FULLTEXT]
18. Van der Putte S, Verhaegen F, Taeymans Y, Thierens H. Correlation of patient skin doses in cardiac interventional radiology with dose-area product. Br J Radiol 2000;73:504-13.  Back to cited text no. 18    
19. Tsapaki V, Kottou S, Vano E, Faulkner K, Giannouleas J, Padovani R, et al . Patient doses in a dedicated Greek cardiac centre. Br J Radiol 2003;76:726-30.  Back to cited text no. 19  [PUBMED]  [FULLTEXT]
20. Hart D, Hillier MC, Wall BF. Doses to patients from medical X ray examinations in the UK. Review, NRPB-W14. HMSO: London; 2000.  Back to cited text no. 20    
21. Kuon E, Glaser C, Dahm JB. Effective techniques for reduction of radiation dosage to patients undergoing invasive cardiac procedures. Br J Radiol 2003;76:406-13.  Back to cited text no. 21  [PUBMED]  [FULLTEXT]
22. Kuon E, Dorn C, Schmitt M, Dahm JB. Radiation dose reduction in invasive cardiology by restriction to adequate instead of optimized picture quality. Health Phys 2003;84:626-31  Back to cited text no. 22  [PUBMED]  [FULLTEXT]
23. Livingstone RS, Raghuram L, Ipeson PK, Raj DV. Evaluation of radiation risk and work practices during cerebral interventions. J Radiol Prot 2003;23:327-36.  Back to cited text no. 23    


    Figures

  [Figure - 1]
 
 
    Tables

  [Table - 1], [Table - 2], [Table - 3]


This article has been cited by
1 Exposure of cardiologists from interventional procedures
G. Kuipers, X. L. Velders, J. J. Piek
Radiation Protection Dosimetry. 2010; 140(3): 259
[VIEW]
2 Impact of biplane versus single-plane imaging on radiation dose, contrast load and procedural time in coronary angioplasty
Sadick, V., Reed, W., Collins, L., Sadick, N., Heard, R., Robinson, J.
British Journal of Radiology. 2010; 83(989): 379-393
[Pubmed]



 

Top
Print this article  Email this article
Previous article Next article

    

 
   Search
 
   Next article
   Previous article 
   Table of Contents
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (93 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
    Materials and Me...
    Results
    Discussion
    Conclusion
    Acknowledgment
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed 2767    
    Printed 99    
    Emailed 0    
    PDF Downloaded 245    
    Comments  [Add]    
    Cited by others  2    

Recommend this journal